Key features
Defibrillator analyzer
Monophasic and biphasic dc
energy measurement
Energy and cardioversion mea-
surement
Peak voltage, peak current, and
overshoot measurement
2-line x 24-character display
Bidirectional RS-232 port for
computer control
Storage and playback of output
waveform so results can be
viewed in greater detail
Optional Ansur test automation
software to standardize testing
procedures, capture waveforms
and test results, and print and
document test results
The QED 6 provides a
scalable solution to accu-
rately test defibrillators.
Lightweight and portable,
the QED 6 measures a
wide range of defibrillator
energy output parameters.
An RS-232 serial
port supports remote
computer control and test
documentation.
Output energy test
Load 50 ± 1 %, with inductance < 70 µ h
Resolution High-range: 1 J
Low-range: 0.1 J
Low-range 0 J to 100 J
High-range 0 J to 400 J
Pulse width 1 ms to 50 ms
Maximum current Low: 35 A
High: 110 A
Maximum voltage Low: 1750 V
High: 5500 V
Minimum voltage Low: 20 V
High: 66 V
Accuracy 400 J range: ± 2 % of reading
100 J to 400 J: ± 2 J
100 J Range: ± 2 % of reading, ± 0.1 J
Waveform storage Discharge viewable via ECG output, paddles, and scope
output
Time expansion lead II
amplitude
High = 3000 V / mV
Low = 900 V / mV
Peak/overshoot
Voltage accuracy 400 J range: ± 10 V
100 J Range: ± 25 V
Current accuracy: ± 1 A
Cardioversion synchronization
test
Measurement from peak or base of simulated R-wave: 0
ms to 199.9 ms
Accuracy: 1 % of full scale or ± 2 ms, whichever is
greater
QED 6
Debrillator Analyzer
Specications