Dynatech MP-7000 Spezifikationen Seite 24

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 23
24
Optional accessories
20200 Ansur Test Software, ESA601
Plug-In
48052FG Banana/ECG adapter
PRINTR/414-US120V Printer, Seiko
DPU-414-30B, with 120 V (2235375)
power supply
(Note: requires additional purchase of
serial printer cable 2461993)
PRINTR414-SHK220V Printer, Seiko
DPU-414-30B, with 220 V (2235382)
power supply
(Note: requires additional purchase of
serial printer cable 2461993)
61096 Printer 120 V power supply
61097 Printer 220 V power supply
97116 DPU-414 and DPU-411 Printer
paper (minimum 7 rolls - price is per
roll)
1800001 North American 220 V
adapter kit
48383FG Lead set, red
48382FG Lead set, black
48201FG Adapter, banana-alligator
75057 Detachable cord set,
240 V/10 A - USA
75033 Detachable cord set,
120 V/15 A - USA
15703 Data transfer cable, RS-232
Ordering information
ISA601PROXL-SHK-P Schuko
receptacle, english overlay, with
internal printer
ISA601PROXL-SHK Schuko receptacle,
english overlay, without internal
printer
ISA601PROXL-UK-P UK receptacle,
english overlay, with internal printer
ISA601PROXL-UK UK receptacle,
english overlay, without internal
printer
ISA601PROXL-AUS-P Australian
receptacle, english overlay, with
internal printer
ISA601PROXL-AUS Australian
receptacle,english overlay, without
internal printer
ISA601PROXL-US-P 601 Pro Series
XL
USP: US receptacle, english overlay,
with internal printer
ISA601PROXL-US US receptacle,
english overlay, without internal
printer
ISA601PROLX-BRAZ-P Brazil
receptacle, english overlay, with
internal printer
ISA601PROXL-BRAZ Brazil receptacle,
english overlay, without internal printer
Included accessories
MANUAL Operator’s Manual
48382FG Black Test-Lead Set
48383FG Red Test-Lead Set
48201FG 5 Banana/Alligator Adapters
Test Receptacle and Cordset
601 Pro Series
XL
Electrical Safety Analyzer
Specications
Electrical Safety Analyzers
IEC60601-1 and AAMI leakage current
Range 0 µA to 8000 µA true-rms or dc only
Accuracy (per IEC60601-1 or AAMI)
DC to 1 kHz ± 1 % of reading ± 1 µA
1 kHz to 100 kHz ± 2 % of reading ± 1 µA
100 kHz to 1 MHz ± 5 % of reading ± 1 µA
Mains on applied part , equivalent device, and equivalent patient l eakage current
tests the following exceptions apply
Applied voltage ≥ 110 % of mains voltage
Accuracy ± 2 % of reading ± 6 µA
IEC61010-1 leakage current
Range 0 µA to 16000 µA true-rms or dc only
Accuracy (per IEC 61010-1 Fig. A.1 filter)
DC to 1 kHz ± 2 % of reading ± 1 µA
1 kHz to 100 kHz ± 4 % of reading ± 2 µA
100 kHz to 1 MHz ± 10 % of reading ± 10 µA
Waveform simulation
Normal sinus rhythm 30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM
Performance pulse 30 BPM and 60 BPM
Sine 10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz
Square 0.125 Hz
2 Hz (50 % duty cycle)
Triangle 2 mV
2 Hz
Arrhythmia A-Fib
A-Flutter
A-TAC
Idioventricular
PVC1
R-on-T
Run
V-Fib
V-Tach
Performance ± 2 % of reading for rate of ± 5 % of reading for amplitude,
fixed at
1 mV peak on a Lead II ECG connection (except for triangle
wave,
which is 2 mV peak to peak)
General
Power Autoswitching, 90 V ac to 265 V ac
Dimensions (LxWxH) 42.2 cm x 30 cm x 14.1 cm (16.62 in x 11.75 in x 5.56 in)
Weight 7.7 kg (17 lb)
riešenia na presné meranie
http://www.elso.sk
Seitenansicht 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 67 68

Kommentare zu diesen Handbüchern

Keine Kommentare