Multi-Function Simulators
ProSim 8
Vital Signs Simulator
Cardiac output
Catheter type Baxter Edwards, 93a-131-7f
Calibration coeffecient 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood temperature 36 °C (98.6 °F) to 38 °C (100.4 °F) ± 0.2 °C in 1 °C steps
Injectate volume 10 cc
Injectate temperature 0 °C or 24 °C
Cardiac output 2.5, 5, 10 liters per minute ± 5 %
Faulty-injectate curve Waveform for simulation available
Left-to-right-shunt curve Waveform for simulation available
Calibrated pulse 1.5 ° for 1 second
Connector Circular DIN 7 pin
Power-on default 5 liters per minute, 0 °C injectate, 37 °C blood temperature
Non-invasive blood pressure
Pressure units mmHg or kPa
Manometer (pressure meter) Range 10 mmHg to 400 mmHg
Resolution 0.1 mmHg
Accuracy ± (0.5 % reading + 0.5 mmHg)
Pressure source Target pressure range 20 mmHg to 400 mmHg
Resolution 1 mmHg
NIBP simulations Pulse 2 mmHg max into 500 ml NIBP system
Volume of air moved 1.25 ml max
Simulations (systolic/diastolic [MAP]) Adult: 60/30 (40), 80/50 (60); 100/65 (77); 120/80
(93); 150/100 (117); and 200/150 (167) and
255/195 (215)
Neonatal: 35/15 (22); 60/30 (40); 80/50 (60); 100/65
(77); 120/80 (93) and 150/100
Pressure variability: systolic and diastolic pressures
are variable by 1 mmHg
NIBP simulations cont. Repeatability Within ± 2 mmHg (at maximum pulse size indepen-
dent of device under test)
Synchronization: normal Sinus heart rates: 30 BPM to
240 BPM
Maximum rate at 1 ml: 240 BPM achievable with
pulses up to 1 ml
Maximum rate at 1.25 ml:
180 BPM
Synchronization: arrhythmias Premature atrial contraction (PAC), premature
ventricular contraction (PVC), atrial fibrillation, and
missed beat
Leak test Target pressure 20 mmHg to 400 mmHg
Elapse time 0:30 min to 5:00 minutes: seconds in 30 second
steps
Leakage rate 0 mmHg/minute to 200 mmHg/minute
Pressure relief test range 100 to 400 mmHg