Dynatech MP-7000 Spezifikationen Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
7
Defibrillator/Transcutaneous Pacer Analyzers
Optional accessories
ANSUR IMPULSE 7000 Ansur Impulse
6000D/7000DP Plug-In
I6/7 D/P ADPT106 MedtronicERS/
Physio-Control (FAST PATCH) (set of
two): 4 mm defibrillator adapters
I6/7 D/P ADPT103 Kimberly Clark/R2
Darox MRL/MDE/NK: 4 mm
defibrillator adapters
I6/7 D/P ADPT102 Internal discharge
paddle contacts (set of two)
I6/7 D/P ADPT105 Medtronic ERS/
Physio-Control (QUIK PACE) (set of
two): 4 mm pacer adapters
I6/7 D/P ADPT110 Zoll Medical NTP/
PD1400: 4 mm pacer adapters
I6/7 D/P ADPT104 Medtronic ERS/
Physio-Control (QUIK COMBO): 4 mm
defib/pacer adapters
I6/7 D/P ADPT107 Philips/Agilent/HP
(CODEMASTER Series-Round):
4 mm defib/pacer adapters
I6/7 D/P ADPT108 Philips/Agilent
HEARTSTART FR2/MRX: 4 mm defib/
pacer adapters
I6/7 D/P ADPT109 Zoll PD-2200
Multi-Function PD-Series, M-Series,
M-Series CCT, AED PRO
®
and AED
Plus defib/pacer adapters
I6/7 D/P ADPT101 GE Marquette
(RESPONDER 1500/1700 Series) (set of
two): 4 mm defib/pacer adapters
IMPULSE 7010 Impulse 7010
Defibrillator Selectable Load Accessory
Transcutaneous Pacemaker Analyzer (Impulse 7000DP only)
Defibrillator input
Fixed load: 50 Ω
Accuracy: ± 1 %, non-inductive (< 2 μH)
Pacemaker input
Variable load: 50 Ω to 1500 Ω by 50 Ω
Accuracy: ± 1 %, non-inductive (< 2 μH)
Manufacturer specific
algorithms
Medtronic/Physio Control LIFEPAK
Philips/Agilent/HP
ZOLL Medical
GE Responder (1500 and 1700)
MRL/Welch Allyn
Schiller Medical
MDE300 (Medical Data Electronics), plus a general purpose
default algorithm selection
Current Range: 4 mA to 250 mA
Accuracy: ± 1 % of reading +0.02 mA
Pulse rate Range: 5 PPM to 800 PPM
Accuracy: ± 0.5 % of reading +0.1 PPM
Pulse width Range: 1 ms to 100 ms
Accuracy: ± 0.5 % of reading +0.01 ms
Demand and asynchronous
mode test
Underdrive rate: 10 BPM minimum
Overdrive rate: 300 BPM maximum
Sensitivity test Automatic interactive threshold detection
Compatible pacer rates: 30 PPM to 120 PPM
ECG R wave
Waveforms: Square, triangle, sine
Widths: 1 ms to 19 ms (by 1 ms), 20 ms to 95 ms (by 5 ms),
100 ms to 300 ms (by 25 ms)
Accuracy: ± 5 % of setting
Amplitude: 0.05 mV to 0.95 mV (by 0.05 mV), 1 mV to 5 mV
(by 0.5 mV)
Accuracy: ± 5 % of setting
Refractory period tests Paced refractory period 20 ms to 500 ms
Sensed refractory period 15 ms to 500 ms
Accuracy: ± 1 ms
General information
Dimensions (LxWxH) 32 cm x 24 cm x 13 cm (13 in x 9.5 in x 5 in)
Weight 3.02 kg (6.6 lb)
Standards
Safety standards CE: IEC/EN61010-1 2nd Edition; Pollution degree 2; CSA:
CAN/CSA-C22.2 NO,61010-1, UL61010-1; C-Tick: Australian
EMC
Discharge curve at 25 Ohms using Ansur and the 7010 load box.
Impulse 6000D/7000DP
Debrillator/Transcutaneous Pacemaker Analyzer
Specications
riešenia na presné meranie