20 Electrical Safety Analyzers
Optional accessories
ANSUR ESA620 Ansur ESA620 Plug-In
Ordering information
ESA620 United States, 115 V, 20 A
ESA620-02 Europe, 230 V
ESA620-01 France, 230 V
ESA620-03 Israel, 230 V
ESA620-05 Australia, 230 V
ESA620-06 United Kingdom, 230 V
ESA620-07 Switzerland, 230 V
ESA620-08 Thailand, 230 V
ESA620-10 230VBRAZ Brazil, 230 V
TA-ESA620-USA ESA620, United
States 115V 20A w/Test Automation
TA-ESA620-EUR ESA620, Europe
230V w/Test Automation
TA-ESA620-FR ESA620, France 230V
w/Test Automation
TA-ESA620-ISR ESA620, Israel 230V
w/Test Automation
TA-ESA620-AUS ESA620, Australia
230V w/Test Automation
TA-ESA620-SWI ESA620, Switzerland
230V w/Test Automation
TA-ESA620-UK ESA620, United
Kingdom 230V w/Test Automation
Included accessories
CD-ROM Operator’s Manual CD
MANUAL Multilingual Getting Started
Guide
ESA620 Accessory Kit
(country specific)
2719-0154 15 A to 20 A adapter
(US only)
SOFT CASE Carry case
Detachable Power Cord (country
specific)
Included accessories for ESA620
with test automation
All of the above, plus:
ANSUR ESA620 Ansur Test-
Automation Software ESA620 plug-in
CABLE ASSEMBLY Data Transfer Cable
Leakage current (continued)
Mains on applied part
test voltage
110 % of mains @ 230 V for IEC 60601
100 % of mains @ 115 V per AAMI
100 % of mains @ 230 V per 62353
Note: For Alternative and Direct applied parts leakage tests, the leak-
age values are compensated for nominal mains as per 62353. Thus,
the accuracy specified for other leakages is not applicable. The actual
leakage readings given during these tests will be higher
Differential leakage
Ranges 50 µA to 199 µA
200 µA to 1999 µA
2 mA to 20 mA
Accuracy ± 10 % of reading ± (2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges
0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ
Accuracy ± (2 % of reading + 2 counts)
± (7.5 % of reading + 2 counts)
Source test voltage 500 V dc
250 V dc
ECG performance waveforms
Accuracy ± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed @ 1 mV
Lead II configuration
Waveforms Rates
ECG complex (BPM): 30, 60, 120, 180, and 240
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz): 0.125 and 2
Sine wave (Hz): 10, 40, 50, 60, and 100
Triangle wave (Hz): 2
Pulse (63 ms pulse width): 30 and 60
Power ratings
Mains voltage outlet 120 V ac
230 V ac
Mains voltage inlet
power range
90 V ac to 132 V ac rms
180 V ac to 264 V ac rms
Maximum current 20 A @ 120 V ac
16 A @ 230 V ac
Hz 50 or 60
Physical case
Dimensions (LxWxH) 31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in)
Weight 4.7 kg (10.25 lb)
Certifications
Certifications CE: IEC/EN61010-1 2nd Edition; Pollution degree 2
CSA: CAN/CSA-C22.2 No 61010-1; UL61010-1
C-Tick: Australian EMC
Environmental
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 90 % non-condensing
Altitude To 2,000 meters @ 230 V ac (main supply voltage)
To 5,000 meters @ 120 V ac (mains supply voltage)
ESA620