30 Non-Invasive Blood Pressure Simulators
Key features
• Dynamic BP simulators for arm-
and wrist-cuff monitors
• Physiological waveform
• Internal pump for high- and
low-pressure release verifica-
tion, leak testing, and pressure
sourcing
• Preset mode for simulation of
most patient conditions
• Five-leads synchronized ECG
simulation
• Arrhythmia simulations,
including premature atrial
contractions #1 and #2, atrial
fibrillation, and PVCs
• Optional Ansur test automation
software to standardize testing
procedures, capture waveforms,
and print and document test
result
Optional accessories
ANSUR BP PUMP 2 Ansur BP Pump 2
Plug-in
75034 Data Transfer Cable
5022010 Soft-sided Vinyl Carrying
Case
2780512FG ECG Adapter Block (allows
simulation of 5-lead ECG waveforms)
PRINTR/414-US120V Printer, Seiko
DPU-414-30 B, 120 V power supply
PRINTR414-SHK220V Printer, Seiko
DPU-414-30B, 200 V power supply
61096 Printer, 120 V power supply
61097 Printer, 220 V power supply
75034 Serial Cable, D9M-D9F
71072 Parallel Printer Cable, D25M-
C36M
5027-0203FG Neonatal/external cuff
mandrel (truncated plastic cylinder
diameters: 7.6, 10, and 14 cm)
98175FG Wrist cuff mandrel (adult)
5215-0269FG Adult Cuff Mandrel
Spacer Block (three required)
5215-0268FG Adult Cuff Mandrel End
Block (two required)
BPPM2M-2001 500 ML Rigid
Aluminum Chamber for NIBP Testing
BPPM2M-2002 100 ML Rigid
Aluminum Chamver for NIBP Testing
700PMP External Pressure Pump
Quick NIBP monitor testing bundles
PS420/DPM1B Bundle PS420/DPM1B
Bundle Kit (includes PS420, DPM1B,
all accessories, and a custom carrying
case)
PS410/DPM1B Bundle PS410/DPM1B
Bundle Kit (includes PS410, DPM1B,
all accessories, and a custom carrying
case)
Pressure generation/
measurement
Static-pressure range: 50 mmHg to 400 mmHg (53 kPa)
Difference between target pressure and actual pressure: -5 mmHg
Internal leak rate: < 2 mmHg per minute with minimum volume of 300 cc
Four respiratory artifacts 3 spontaneous breathing; controlled ventilation
Three adult wrist-cuff simu-
lations
Normal, Hyper, Hypo
Pressure source Specified pressure generated from 50 mmHg to 400 mmHg in selectable incre-
ments of 1 mmHg
Pressure gauge Static pressure measured from 0 mmHg to 400 mmHg at the pressure port
Pressure relief test Test for the NIBPM pressure relief valve (0 mmHg to 400 mmHg) with display of
peak pressure
Neonate internal cuff simu-
lations
Internal neonate cuff; four standard neonate pressures
Neonate simulations Cuff #1: Blood pressure: 35/15 Heart rate: 120 BPM Pulse volume: 0.1 cc
Cuff #2: Blood pressure: 60/30 Heart rate: 120 BPM Pulse volume: 0.1 cc
Cuff #3: Blood pressure: 80/50 Heart rate: 120 BPM Pulse volume: 0.1 cc
Cuff #4: Blood pressure: 100/70 Heart rate: 120 BPM Pulse volume: 0.1 cc
Irregular pulse BP and ECG: Premature atrial contractions # 1, premature atrial contractions # 2,
premature ventricular contractions, atrial fibrillation and PVCs
User-definable simulations User-definable systolic and diastolic values, along with heart rate and pulse volume
Ranges:
Systolic pressure Diastolic pressure Heart rate Pulse volume
range range
20 mmHG to 10 mmHG to 30 BPM to 0.1 cc to 2.4 cc in
250 mmHG 200 mmHG 250 BPM increments of 0.1 cc
Simulation parameters
performance
Max pulse volume: 2.4 cc
Max heart rate: 200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume
Internal neonatal cuff volume: 20 cc
Internal adult cuff volume (including NN volume): 310 cc
Heart rate setting accuracy: ± 1 BPM
Simulation units: kPa and mmHg (user selectable)
Pressure leak test The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the
pressure loss over time. Peak pressure and present pressure are displayed at all
times; leak rate is displayed when it is available.
Autosequences Nine autosequences are provided for four tests and up to five simulations
Electrical ECG (optional) Signals: RA, LA, RL, LL, V
Waveform: Lead II
Amplitude: 1 mV peak (± 10 %) NIBP peripheral pulse synchronized with ECG signal
Connections: Optional external ECG adapter, physiological synchronization with
NIBP
The BP Pump 2 is a second-
generation non-invasive
blood pressure (NIBP) moni-
tor analyzer that efficiently
verifies oscillometric adult and
neonatal NIBP. The BP Pump
2’s unique feature set includes
tests to accurately interrogate
wrist-cuff monitors, internal
cuff volumes, and optional
5-lead synchronized ECG sim-
ulations for spot checks on the
monitor. The simulated periph-
eral pulse is synchronized with
this electrical ECG signal for
testing NIBP monitors utilizing
gated measurement for noise/
artifact rejection.
BP Pump 2
Non-Invasive Blood Pressure Simulator
Specications