CHAPTER 4LIFE CYCLE COUNTERS
70 9650-0801-01 Rev. A
CO2
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.
SpO2
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>SpO2>Error Counts.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.
Defibrillator/Pacer
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>Defib/Pacer>Error
Counts.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.