Service Manual Overview 7
Identifying Propaq Encore Configurations
The following table identifies Propaq monitor configurations and how they are indicated.
The model-option number and serial number are located on the back of the housing. The
monitor indicators are located under the handle on the back.
WARNING If the monitor is opened for repair or calibration, a dielectric strength
test must be completed to ensure the integrity of the patient isolation barrier. The
dielectric strength test procedure is provided in “Functional Verificationon
page 11.
Table 1. Monitor Configurations
Product Model Option Parameters
a
a. The Acuity option is available with all models.)
Propaq Encore Monitor 202-EL ECG, NIBP, Two Temps, Resp, EL Display
204-EL ECG, NIBP, Two Temps, IBP, Resp, EL Display
206-EL ECG, NIBP, Two Temps, 2 IBP, Resp, EL Display
SpO2-Only Module 210 SpO2 and one of the following:
Nellcor MP-405 (Top level assembly 007-0059-XX)
Nellcor MP-203 (Top level assembly 007-0020-XX)
Nellcor MP-507 (Top level assembly 007-0109-XX)
Masimo MS3 (Top level assembly 007-0060-XX)
Expansion Module (See next table.)